Specifications

RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DM3000 Series
2-66
Return Format:
:trigger:vmcomplete:polar? returns the current VMC output polar.
9. :trigger:vmcomplete:pulsewidth
Syntax:
:trigger:vmcomplete:pulsewidth?
:trigger:vmcomplete:pulsewidth
<value>
Function:
The command queries and sets VMC output pulsewidth.
Parameters:
Name
Type
Range
Default
<value>
consecutive integer
1~30ms
1~200ms
1~400ms
30ms
100ms
100ms
Explanations:
:trigger:vmcomplete:pulsewidth? queries the VMC output pulsewidth.
:trigger:vmcomplete:pulsewidth
<value>
sets the
VMC output pulsewidth. The
range of value: the resolution is , 1~30ms, and the default is 30ms; the resolution
is , 1~200ms, and the default is 100ms; the resolution is 6½ , 1~400ms, and the
default is 100ms.
Remarks: For ACV and ACI measurement, the range of value: the resolution is ,
1~30ms, and the default is 30ms; the resolution is 4½ , 1~200ms, and the default is
100ms; the resolution is 5½ , 1~400ms, and the default is 100ms.
Return Format:
:trigger:vmcomplete:pulsewidth? returns the current VMC output pulsewidth.