Specifications

RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DM3000 Series
2-41
6. :resolution:resistance
Syntax:
:resolution:resistance?
:resolution:resistance{
<range>
|MIN|MAX|DEF}
Function:
The command queries and sets the measure precision of two circuits resistance.
Parameters:
Name
Range
Default
<range>
{0|1|2}
2
Explanations:
:resolution:resistance? queries the measure precision of the two circuits resistance.
Before sending the command, please turn on two circuit resistance measure function.
:resolution:resistance{
<range>
|MIN|MAX|DEF} sets the measure precision of the
two circuits resistance.
The Table 2- 8 indicates the parameters and their meanings.
Return Format:
:resolution:resistance? returns the current parameter: 01 or 2.(see Table 2- 8)
Example:
To set the measure precision of the two circuits resistance as 5 ½ :
:resolution:resistance1
7. :resolution:fresistance
Syntax:
:resolution:fresistance?
:resolution:fresistance{
<range>
|MIN|MAX|DEF}
Function:
The command queries and sets the measure precision of four circuits resistance.