9. :measure:voltage:DC:ratio:digit
Syntax:
:measure:voltage:DC:ratio:digit?
:measure:voltage:DC:ratio:digit□{INC|DEC|5|6|7}
Function:
The command queryies and sets the digit of the ratio of DC voltages in two circuits.
Explanations:
:measure:voltage:DC:ratio:digit? queries the digit of the ratio of the DC voltages in
two circuits.
:measure:voltage:DC:ratio:digit□
<value>
sets the digit of the ratio of the DC
voltages in two circuits.
The Table 2- 2 indicates the parameters and their meanings.
Remarks: If the showing digit reachs 5, DEC command will be invalid. If 7,INC
invalid.
Return Format:
:measure:voltage:DC:ratio:digit? returns the current digit: 5、6 or 7.
Examples:
To set the digit of the ratio as 6:
:measure:voltage:DC:ratio:digit□6
To increase the digit to 7:
:measure:voltage:DC:ratio:digit□INC
10. :measure:voltage:AC?
Syntax:
:measure:voltage:AC?
Function:
The command queries AC voltage.
Return Format:
The query returns the current AC voltage in the form of 3.941713e-01, unit is V(for