Specifications

RIGOL
© 2008 RIGOL Technologies, Inc.
Programming Guide for DM3000 Series
1-5
Commands Introduction
To meet the needs of different users, DM3000 provides RIGOL commands and
another two commands that are compatible with the related products.
RIGOL DM3000 commands
The commands compatible with Agilent
The commands compatible with Fluke
Send CMDSET command to change the commands type. For the usage please see
the follows:
Syntax:
CMDSET{RIGOL|AGILENT|FULUKE}
CMDSET
Function:
The command sets and queries command types.
The first command parameters include RIGOLAGILENT and FLUKE, and it means
choosing the corresponding commands.
The second command queries the current commands, and returns the
commands name.
Parameters:
Name
Range of values
Default
Digital parameter
{RIGOL|AGILENT|FLUKE}
RIGOL
Remarks
If you dont want to send this commands when strating machine, the default will be
ROGOL.