User manual
Chapter 2 Command System RIGOL
DG800 Programming Guide 2-25
*PSC
Syntax
*PSC {0|1}
*PSC?
Description
Enables or disables clearing of the status byte enable register and standard event enable register at
power-on.
Queries whether to clear the status byte enable register and standard event enable register at power-on.
Parameter
Name
Type
Range
Default
{0|1}
Discrete
0|1
1
Remarks
The *PSC 1 command means clearing the status byte enable register and standard event enable
register at power-on. The *PSC 0 command means the status byte enable register and standard event
enable register will not be affected at power-on.
You can also send the *SRE
0 and *ESE 0 commands to clear the status byte enable register and
standard event enable register respectively.
Return Format
The query returns 0 or 1.
Example
*PSC 1 /*Enables clearing of the status byte enable register and standard event enable register at
power-on.*/
*PSC? /*Queries whether to enable or disable the clearing of the affected registers at power-on. The
query returns 1.*/
*RCL
Syntax
*RCL {0|1|2|3|4|5}
Description
Recalls the state file stored in the specified location of the internal non-volatile memory.
Parameter
Name
Type
Range
Default
{0|1|2|3|4|5}
Discrete
0|1|2|3|4|5
None
Remarks
The instrument provides 6 storage locations (No. 0 to 5) in the internal memory for storing the state
files. Sending this command can recall the state file stored in the specified storage location of the
internal non-volatile memory. Select the files from No. 0 to No. 5 to recall the state file stored in the
corresponding storage location respectively.
This command is only valid when an effective state file is found in the specified storage location of the
internal non-volatile memory.
The state file stored includes the dual-channel selected waveform, frequency, amplitude, offset, duty
cycle, symmetry, phase, modulation, sweep, burst parameters, frequency counter parameters, as well
as utility function parameters and system parameters under the Utility menu.