User Guide

RIGOL Chapter 2 Basic Waveform Output
DG4000 Series Users Guide
2-18
Basic Waveform Output Example
Configure the generator to output a pulse waveform with 1.5MHz frequency,
500mVpp amplitude, 5mV
DC
DC offset, 200ns pulse width, 75ns leading (rising)
edge time, 100ns trailing (falling) edge time and 5ns delay.
1. Press CH1 at the front panel and the backlight turns on, indicating that CH1 is
selected.
2. Press Pulse at the front panel and the backlight turns on, indicating that
Pulse waveform is selected.
3. Press Freq/Period to highlight Freq. The lightspot above the number
indicates current cursor location (as shown in the figure below). Use the
numeric keyboard or direction keys and knob to input the frequency value
1.5 and select the unit MHz from the pop-up menu.
Figure 2-5 To Set Waveform Parameters
4. Press Ampl/HiLevel to highlight Ampl. The lightspot above the number
indicates current cursor location. Use the numeric keyboard or direction keys
and knob to input the amplitude value 500 and select the unit mVpp from
the pop-up menu.