User Guide

RIGOL Chapter 6 Sweep
DG4000 Series Users Guide
6-4
If the center frequency is 550Hz, the frequency span range is ± (550Hz1μHz)
= ± 1.099999998kHz; if the center frequency is 155MHz, the frequency span range
is ± (160MHz155MHz) = ± 10MHz.
The generator will renewedly sweep from the specified Start Frequency after
Center Frequency or Frequency Span is changed.
Tip
In large-scale sweep, the amplitude characteristic of the output signal might
change.
Sweep Type
DG4000 provides three sweep types, including Linear, Log and Step, and the
default is Linear.
Linear Sweep
The output frequency of the instrument varies linearly in the way of several Hertz
per second. The variation is controlled by Start Frequency, End Frequency and
Sweep Time.
When Sweep is enabled, press SwpType to select Linear. A line is displayed on
the waveform on the screen, indicating that the output frequency changes in linear
mode.