Instructions
Chapter 2 Command System  RIGOL 
DG2000 Programming Guide  2-7 
:COUN?       /*Queries the status of the frequency counter and the query returns OFF.*/ 
:COUN 1 /*Enables the frequency counter function.*/ 
:COUN? /*Queries the status of the frequency counter and the query returns RUN (the default 
operating status).*/ 
:COUN STOP    /*Sets the operating status of the frequency counter to "STOP".*/ 
:COUN?         /*Queries the status of the frequency counter and the query returns STOP.*/ 
:COUNter:STATIstics:CLEAr 
Syntax 
:COUNter:STATIstics:CLEAr 
Description 
Clears the statistics results. 
Remarks   
  This command is only valid when the statistical function of the frequency counter is enabled 
(
:COUNter:STATIstics[:STATe]). 
  The statistics results are cleared automatically when the statistical function of the frequency counter is 
disabled. 
:COUNter:STATIstics[:STATe] 
Syntax 
:COUNter:STATIstics[:STATe] {ON|1|OFF|0} 
:COUNter:STATIstics[:STATe]? 
Description 
Enables or disables the statistical function of the measurement results of the frequency counter. 
Queries the on/off status of the statistical function of the measurement results of the frequency counter. 
Parameter 
Name 
Type 
Range 
Default 
{ON|1|OFF|0} 
Bool 
ON|1|OFF|0 
OFF 
Return Format 
The query returns ON or OFF. 
Example   
:COUN:STATI ON  /*Enables the statistical function of the measurement results of the frequency 
counter.*/ 
:COUN:STATI?      /*Queries the on/off status of the statistical function of the measurement results of the 
frequency counter and the query returns ON.*/ 