Instructions
Chapter 2 Command System  RIGOL 
DG2000 Programming Guide  2-129 
Return Format 
The query returns the return time in scientific notation, with 7 effective digits. For example, 1.000000E+00 
(the return time is 1 s). 
Example 
:SOUR1:SWE:RTIM 1    /*Sets the return time of the sweep function of CH1 to 1 s.*/ 
:SOUR1:SWE:RTIM?     /*Queries the return time of the sweep function of CH1 and the query returns 
1.000000E+00.*/ 
[:SOURce[<n>]]:SWEep:SPACing 
Syntax 
[:SOURce[<n>]]:SWEep:SPACing {LINear|LOGarithmic|STEp} 
[:SOURce[<n>]]:SWEep:SPACing? 
Description 
Sets the sweep type of the specified channel to Linear (LINear), Log (LOGarithmic), or Step (STEp). 
Queries the sweep type of the specified channel. 
Parameter 
Name 
Type 
Range 
Default 
[<n>] 
Discrete 
1|2 
1 
{LINear|LOGarithmic|STEp} 
Discrete 
LINear|LOGarithmic|STEp 
LINear 
Remarks   
  DG2000 provides Linear, Log, and Step sweep types. 
  In the linear sweep type, the output frequency of the instrument varies linearly in the way of "several 
Hertz per second". The variation is controlled by "Start Frequency" 
(
[:SOURce[<n>]]:FREQuency:STARt), "Stop Frequency" ([:SOURce[<n>]]:FREQuency:STOP), and 
"Sweep Time" (
[:SOURce[<n>]]:SWEep:TIME). A straight line is displayed on the waveform on the 
screen, indicating that the output frequency varies linearly. 
  In Log Sweep type, the output frequency of the instrument varies in a logarithmic fashion, that is, the 
output frequency changes in the way of "octave per second" or "decade per second". The variation is 
controlled by "Start Frequency", "Stop Frequency", and "Sweep Time". An exponential function curve 
is displayed on the screen, indicating that the output frequency changes in a logarithmic mode. 
  In Step Sweep type, the output frequency has a stepwise increase from "Start Frequency" to "Stop 
Frequency". The dwell time that the output signal stays at each frequency point is determined by 
"Sweep Time" and "Steps". A step waveform is displayed on the screen, indicating that the output 
frequency varies with a stepwise increase. 
  When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related parameters of CH1 by default. 
Return Format 
The query returns LIN, LOG, or STE. 
Example 
:SOUR1:SWE:SPAC LIN    /*Sets the sweep type of CH1 to Linear.*/ 
:SOUR1:SWE:SPAC?          /*Queries the sweep type of CH1 and the query returns LIN.*/ 