Instructions
RIGOL  Chapter 2 Command System 
2-94  DG2000 Programming Guide 
:SOURce[:MOD]:AM 
Command List: 
  [:SOURce[<n>]][:MOD]:AM[:DEPTh] 
  [:SOURce[<n>]][:MOD]:AM:DSSC 
  [:SOURce[<n>]][:MOD]:AM:INTernal:FREQuency 
  [:SOURce[<n>]][:MOD]:AM:INTernal:FUNCtion 
  [:SOURce[<n>]][:MOD]:AM:SOURce 
  [:SOURce[<n>]][:MOD]:AM:STATe 
[:SOURce[<n>]][:MOD]:AM[:DEPTh] 
Syntax 
[:SOURce[<n>]][:MOD]:AM[:DEPTh] {<depth>|MINimum|MAXimum} 
[:SOURce[<n>]][:MOD]:AM[:DEPTh]? [MINimum|MAXimum] 
Description 
Sets the AM modulation depth of the specified channel. 
Queries the AM modulation depth of the specified channel. 
Parameter 
Name 
Type 
Range 
Default 
[<n>] 
Discrete 
1|2 
1 
<depth> 
Real 
0% to 120% 
100% 
Remarks 
  When [:SOURce[<n>]] or [<n>] is omitted, the system sets the related parameters of CH1 by default. 
  Modulation depth indicates the amplitude variation. It is expressed in percentage. When the 
modulation depth is 0%, the output amplitude is half of the amplitude of the carrier signal. At 100% 
depth, the output amplitude is equal to carrier waveform amplitude. When the modulation depth is 
greater than 100%, the output amplitude of the instrument will not exceed 10 Vpp (when the load is 
50 Ω). 
  When the external modulation source (
[:SOURce[<n>]][:MOD]:AM:SOURce) is selected, the output 
amplitude of the instrument is controlled by the ±5 V signal level on the rear-panel [Sync/Ext 
Mod/Trig/FSK] connector. For example, if the modulation depth is set to 100%, the output 
amplitude will be the maximum when the modulating signal is +5 V and the minimum when the 
modulating signal is -5 V. 
Return Format 
The query returns the AM modulation depth in scientific notation, with 7 effective digits. For example, 
5.000000E+01 (the AM modulation depth is 50°). 
Example 
:SOUR1:AM 50    /*Sets the AM modulation depth of CH1 to 50%.*/ 
:SOUR1:AM?      /*Queries the AM modulation depth of CH1 and the query returns 5.000000E+01.*/ 