Instructions
Appendix    RIGOL 
DG2000 Performance Verification Guide 
3 
Harmonic Distortion Test 
Frequency 
Setting 
Value 
Setting 
Measurement 
Value 
Calculation 
Result
[1]
Specification  Pass/Fail 
10MHz 
Waveform: 
Sine 
Amplitude: 
0dBm 
Offset: 0V
dc
Base waveform: 
  <-55dBc   
2
nd
 order 
harmonic: 
20MHz 
Base waveform:   
<-50dBc   
2
nd
 order 
harmonic: 
40MHz 
Base waveform: 
  <-40dBc   
2
nd
 order 
harmonic: 
60MHz
[2]
Base waveform: 
  <-35dBc   
2
nd
 order 
harmonic: 
Note
[1]
: Calculation result = 2th order harmonic measurement value - base waveform measurement value. 
Note
[2]
: Only applicable to DG2102 and DG2072. 
Spurious Signal Test 
Output 
Frequency 
Start 
Frequency 
Stop 
Frequency 
A  A-0dBm  Specification  Pass/Fail 
5MHz  0Hz  30MHz      <-60dBc   
10MHz  0Hz  50MHz      <-60dBc   
20MHz  0Hz  100MHz      <-54dBc   
30MHz  0Hz  150MHz      <-48dBc   
60MHz
[1] 
0Hz  300MHz      <-30dBc   
Note
[1]
: Only applicable to DG2102 and DG2072. 
Rise/Fall Time Test 
Waveform  Setting  Measurement Value  Specification  Pass/Fail 
Square 
Frequency: 1kHz 
Amplitude: 1Vpp 
Offset: 0V
dc
Rise Time   
Typical (1 Vpp, 1 
kHz) 
≤9 ns 
Fall Time   