RIGOL  Chapter 2 Performance Verification Test 
DG2000 Performance Verification Guide 
2-16 
record the measurement result as A. Calculate the non-harmonic spurious signal 
(A-0dBm) and judge whether it is within the specification range.  
6.  Keep the output impedance of CH1 of DG2000 at 50Ω. Set the output waveform 
of CH1 as a sine waveform with 10MHz frequency, 0dBm amplitude and 0V
dc
offset.  
7.  Keep the reference level, input attenuation, resolution bandwidth, peak offset 
and sweep mode of the spectrum analyzer as 10dBm, 20dB, 1kHz, 3dB and 
single respectively. Set its start frequency to 0Hz and stop frequency to 50MHz.  
8.  Press Sweep/Trig  Single to perform a sweep.  
9.  After the spectrum analyzer finishes a sweep, use Peak and the marker 
function to measure the maximum spurious signal (except harmonics) and 
record the measurement result as A. Calculate the non-harmonic spurious signal 
(A-0dBm) and judge whether it is within the specification range.  
10.  Keep the output impedance of CH1 of DG2000 at 50Ω. Set the output waveform 
of CH1 as a sine waveform with 20MHz frequency, 0dBm amplitude and 0V
dc
offset.  
11.  Keep the reference level, input attenuation, resolution bandwidth, peak offset 
and sweep mode of the spectrum analyzer as 10dBm, 20dB, 1kHz, 3dB and 
single respectively. Set its start frequency to 0Hz and stop frequency to 100MHz.  
12.  Repeat Step 8 and 9. 
13.  Keep the output impedance of CH1 of DG2000 at 50Ω. Set the output waveform 
of CH1 as a sine waveform with 30MHz frequency, 0dBm amplitude and 0V
dc
offset.  
14.  Keep the reference level, input attenuation, resolution bandwidth, peak offset 
and sweep mode of the spectrum analyzer as 10dBm, 20dB, 1kHz, 3dB and 
single respectively. Set its start frequency to 0Hz and stop frequency to 150MHz.  
15.  Repeat Step 8 and 9. 
16.  Keep the output impedance of CH1 of DG2000 at 50Ω. Set the output waveform 
of CH1 as a sine waveform with 60MHz frequency, 0dBm amplitude and 0V
dc
offset.  
17.  Keep the reference level, input attenuation, resolution bandwidth, peak offset 
and sweep mode of the spectrum analyzer as 10dBm, 20dB, 1kHz, 3dB and 
single respectively. Set its start frequency to 0Hz and stop frequency to 300MHz.  