User`s guide

RIGOL
© 2008 RIGOL Technologies, Inc.
Users Guide for DG10× 2 Series
2-12
To Set Pulse Waveform
Press the Pulse button, in the Normal Mode, the operation menu will appear at the
bottom of the screen. Set the Pulse parameters by using the operation menu.
The parameters for Pulse waveforms are:
Frequency/ Period, Amplitude/ High Level,
Offset/ Low Level, Pulse Width/ Duty Cycle and Delay
. See Figure 2-13, in the
operation menu, select Width, and the corresponding parameter will be displayed in
inverse color for which users can make a change.
Figure 2-13
Pulse Signal Parameter Setting Interface
Table 2-4 Pulse Signal Operation Menu
Term Explanation:
Pulse Width: The time span between thresholds of 50% of the rising edge
amplitude to the next 50% of the falling edge amplitude.
Menu
Settings
Explanation
Frequency/
Period
Setting the signal’s frequency or period; the
current parameter will switch at a second press.
Amplitude/
High Level
Setting the signal’s Amplitude or High Level; the
current parameter will switch at a second press.
Offset/Low
Level
Setting the signal’s Offset or Low Level; the
current parameter will switch at a second press.
Width/
DtyCyc
Setting the Pulse Width or Duty Cycle for Pulse
Waveform.
Delay
Setting the Delay time.