User`s guide
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DG10× 2 Series
4-12
In the N-Cycle mode, the Start Phase upper limit is 360°, if the specified phase is
greater than 360°, the system will adjust the Start Phase to 360° automatically.
67. Start Phase lower limit=-360°
In the N-Cycle mode, the Start Phase upper limit is -360°, if the specified phase is
less than -360°, the system will adjust the Start Phase to -360° automatically.
68. Trigger Period upper limit=500s
In the N-Cycle mode, the Trigger Period upper limit is 500s. If the specified
Trigger Period is greater than 500s, the system will adjust the Trigger Period to
500s automatically.
69. Trigger Period lower limit=1μs
In the N-Cycle mode, the Trigger Period lower limit is 1.0μs. If the specified
Trigger Period is less than 1.0μs, the system will adjust the Trigger Period to 1.0μs
automatically.
70. Delay upper limit=40s
In the N-Cycle mode, the Delay upper limit is 40.0s. If the specified Delay is
greater than 40.0s, the system will adjust the Delay to 40.0s automatically.
71. Delay lower limit=0s
In the N-Cycle mode, the Delay lower limit is 0s. If the specified Delay is less than
0s, the system will adjust the Delay to 0s automatically.
72. Start Phase upper limit=+360°
In the infinite mode, the Start Phase upper limit is 360.0
o
. If the specified Start
Phase is greater than 360.0
o
, the system will adjust the Start Phase to 360.0
o
automatically.
73. Start Phase lower limit=-360°
In the infinite mode, the Start Phase lower limit is -360.0
o
. If the specified Start
Phase is less than -360.0
o
, the system will adjust the Start Phase to -360.0
o
automatically.
74. Delay upper limit=40s
In the infinite mode, the Delay time upper limit is 40s. If the specified time is
greater than 40s, the system will adjust the Delay time to 40s automatically.
75. Delay lower limit=0s
In the infinite mode, the Delay time lower limit is 0s. If the specified time is less
than 0s, the system will adjust the Delay time to 0s automatically.
76. Offset upper limit=10V