User`s guide
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DG10× 2 Series
3-22
Example for Channel Copy
Copy the parameters from CH1 to CH2.
The operation steps:
1. Set the parameters of CH1 as shown in the following figure: Sine wave, frequency
1kHz, amplitude 5V
PP
, Offset 0 V
DC
, phase 10°.
Figure 3-22
Parameters Setting in CH1
2. Set the parameters of CH2 as shown in the following figure: Ramp wave,
frequency 1.5 kHz, amplitude 5V
PP
, Offset 0 V
DC
, phase 0°.
Figure 3-23
Parameters Setting in CH2
3. Press Utility Coupling Switch “Coupling Off” CopyCH 1 2
Confirm . Parameters in CH1 will be copied to CH2 as shown in the
following.
Figure 3-24
Parameters in CH2 after the copying
NOTE: Channel copy is limited by parameters inspection. See the explanation in
section “Channel Copy” in page 2-58.