Specifications
RIGOL DG10X2 Commands System
Programming Guide for DG10X2
2-62
Command
Format
COUPling:BASEdchannel?
Function
Query the base channe that has been selected.
Return Value
The query returns CH1 or CH2.
5. COUPling:PHASEDEViation
Command
Format
COUPling:PHASEDEViation <value>
Function
Set the phase deviation, the unit is degree.
Explanations
<value>: -180°~180°
Example
COUP:PHASEDEV 10
6. COUPling:PHASEDEViation?
Command
Format
COUPling:PHASEDEViation?
Function
Query the phase deviation.
Return Value
The query returns the phase deviation in the form of scientific
notation, such as: 1.000000e+01.
7. COUPling:FREQDEViation
Command
Format
COUPling:FREQDEViation <value>
Function
Set the frequency deviation in Hz.
Explanations
<value>: 0Hz~20MHz
Example
COUP:FREQDEV 100
8. COUPling:FREQDEViation?
Command
Format
COUPling:FREQDEViation?
Function
Query the frequency deviation.
Return Value
The query returns the frequency deviation in the form of scientific
notation, such as: 1.000000e+02.
9. COUPling:CHANNCopy
Command
Format
COUPling:CHANNCopy {1>2|2>1}
Function
Copy CH1 to CH2 or copy CH2 to CH1.