Specifications
RIGOL DG10X2 Commands System
Programming Guide for DG10X2
2-44
TRIGger
TRIGger commands are available in Sweep and Burst mode for CH1 only.
DG10X2 supports following TRIGger commands:
1. TRIGger:SOURce
2. TRIGger:SOURce?
3. TRIGger:SLOPe
4. TRIGger:SLOPe?
5. TRIGger:DELay
6. TRIGger:DELay?
The detailed information of each command are:
1. TRIGger:SOURce
Command
Format
TRIGger:SOURce {IMMediate|EXTernal|BUS}
Function
Select the trigger source for generator, such as: internal trigger
(IMM), external trigger (EXT) from the [Ext Trig/FSK/Burst]
connector on the rear panel, or manual trigger (BUS). The default
is IMM.
Example
TRIG:SOUR EXT
2. TRIGger:SOURce?
Command
Format
TRIGger:SOURce?
Function
Query the trigger source of generator.
Return Value
The query returns IMM, EXT or BUS.
3. TRIGger:SLOPe
Command
Format
TRIGger:SLOPe {POSitive|NEGative}
Function
Select whether the generator uses rising edge(POS) or falling
edge(NEG) of the trigger signal inputted from the [Ext
Trig/FSK/Burst] connector on the rear panel. The default is POS
(rising edge).