Specifications

DG10X2 Commands System RIGOL
Programming Guide for DG10X2
2-31
Format
{SINusoid|SQUare|RAMP|NRAMp|TRIangle|NOISe|USER}
Function
Select the internal modulation source of AM.
Explanations
In internal modulation source mode, the modulation wave could be
sine, square, ramp, negative ramp, triangle, noise or arbitrary
wave, the default is sine.
Example
AM:INT:FUNC SQU
4. AM:INTernal:FUNCtion?
Command
Format
AM:INTernal:FUNCtion?
Function
Query the internal modulation wave of AM that has been selected.
Return Value
The query returns SIN, SQU, RAMP, NRAM, TRI, NOIS or USER.
5. AM:INTernal:FREQuency
Command
Format
AM:INTernal:FREQuency {<frequency>|MINimum|MAXimum}
Function
Set the frequency of internal modulation of AM in Hz.
Explanations
Frequency range: 2mHz 20kHz
Example
AM:INT:FREQ 200
6. AM:INTernal:FREQuency?
Command
Format
AM:INTernal:FREQuency?
Function
Query the frequency of internal modulation of AM.
Return Value
The query returns the percent of the frequency of AM internal
modulation in the form of scientific notation, such as:
2.000000e+02.
7. AM:DEPTh
Command
Format
AM:DEPTh {<depth percent>|MINimum|MAXimum}
Function
Set the depth of internal modulation of AM in percent.
Explanations
Depth range: 0% 120%
Example
AM:DEPT 70
8. AM:DEPTh?
Command
AM:DEPTh? [MINimum|MAXimum]