Specifications

DG10X2 Commands System RIGOL
Programming Guide for DG10X2
2-23
Format
Function
Query the state of the [Output] connector on the front panel.
Return Value
The query returns OFF or ON.
3. OUTPut:LOAD
Command
Format
OUTPut:LOAD {<ohm>|INFinity|MINimum|MAXimum}
Function
Select the desired output termination of CH1. The specified value
is only used for amplitude and offset voltage.
Explanations
Ω is the unit of <ohm>, the default is 50Ω.
INFinity sets the output terminal as High Z.
Example
OUTP:LOAD 50
4. OUTPut:LOAD?
Command
Format
OUTPut:LOAD? [MINimum|MAXimum]
Function
Query the current load settings of CH1.
Return Value
The query returns the current load setting in ohms or returns
Infinity.
5. OUTPut:POLarity
Command
Format
OUTPut:POLarity {NORMal|INVerted}
Function
Set the polarity of waveform for CH1.
Example
OUTP:POL NORM
6. OUTPut:POLarity?
Command
Format
OUTPut:POLarity?
Function
Query the polarity of waveform from CH1.
Return Value
The query returns NORM or INV.
7. OUTPut:SYNC
Command
Format
OUTPut:SYNC {OFF|ON}
Function
Disable or enable the rear panel [Output] connector of CH1. The
default is “OFF”.