Specifications

RIGOL DG10X2 Commands System
Programming Guide for DG10X2
2-10
Function
Query the output function from CH1.
Explanations
The query always returns CH1:ARB after sending FUNC DC or FUNC
USER.
Example
The query returns CH1:SIN, CH1:SQU, CH1:RAMP, CH1:PULS,
CH1:NOIS or CH1:ARB, the default is CH1:SIN.
3. FUNCtion:USER
Command
Format
FUNCtion:USER {<name of arbitrary wave>|VOLATILE}
Function
Separately select any one wave from built-in arbitrary waves and 10
user-defined waves for CH1 or select the wave that has been
downloaded into volatile memory.
Explanations
The built-in waves contains:
Common:
NegRamp/AttALT/AmpALT/StairDown/StairUp/StairUD/Cpulse/
PPulse/NPulse/Trapezia/RoundHalf/AbsSine/AbsSineHalf/
SINE_TRA/SINE_VER
Math:
Exp_Rise/Exp_Fall/Tan/Cot/Sqrt/X 2/Sinc/Gauss/HaverSine/
Lorentz/Dirichlet/GaussPulse/Airy
Project:
Cardiac/Quake/Gamma/Voice/TV/Combin/BandLimited/
Stepresponse/Butterworth/Chebyshev1/ Chebyshev2
Window Function:
Boxcar/Barlett/triang/Blackman/Hamming/Hanning/Kaiser
Others:
Roundpm/DC
Send FUNC DC command when use DC.
Abbreviation is invalid.
Example
FUNC:USER VOLATILE
4. FUNCtion:USER?
Command
Format
FUNCtion:USER?
Function
Query the name of arbitrary wave generated from CH1.
Return
Value
The query returns the name of built-in arbitrary wave that has been
selected. (such as EXP_RISE), VOLATILE or returns any name of