Data Sheet
3
RIGOL
Model DG1032Z DG1062Z
Channel 2 2
Max Frequency 30 MHz 60 MHz
Sample Rate 200 MSa/s
Waveform
Basic Waveform Sine, Square, Ramp, Pulse, Noise
Built-in Arbitrary Waveform
160 kinds, including Sinc, Exponential Rise, Exponential Fall, ECG, Gauss, HaverSine, Lorentz,
Dual-Tone, etc.
Frequency Characteristics
Sine 1 μHz to 30 MHz 1 μHz to 60MHz
Square 1 μHz to 15 MHz 1 μHz to 25 MHz
Ramp 1 μHz to 500kHz 1 μHz to 1MHz
Pulse 1 μHz to 15 MHz 1 μHz to 25 MHz
Harmonic 1uHz to 10MHz 1uHz to 20MHz
Noise (-3dB) 30 MHz bandwidth 60 MHz bandwidth
Arbitrary Waveform 1 μHz to 10 MHz 1 μHz to 20 MHz
Resolution 1 μHz
Accuracy ±1 ppm of the setting value, 18
℃
to 28
℃
Sine Wave Spectrum Purity
Harmonic Distortion
Typical (0 dBm)
DC-10 MHz (included): <-65 dBc
10 MHz to 30 MHz (included): <-55 dBc
30 MHz to 60 MHz (included): <-50 dBc
Total Harmonic Distortion <0.075% (10 Hz to 20 kHz, 0 dBm)
Spurious (non-harmonic)
Typical (0 dBm)
≤10 MHz <-70 dBc
>10 MHz <-70 dBc + 6 dB/octave
Phase Noise
Typical (0 dBm, 10 kHz offset)
10 MHz: <-125 dBc/Hz
Signal Characteristics
Square
Rise/Fall Time
Typical (1 Vpp)
<10ns
Overshoot
Typical (100 kHz, 1 Vpp)
≤5%
Duty Cycle 0.01% to 99.99% (limited by the current frequency setting)
Non-symmetry 1% of the period + 5 ns
Jitter (rms)
Typical (1 Vpp)
≤5 MHz 2 ppm + 200 ps
> 5 MHz 200 ps
Ramp
Linearity ≤1% of peak output (typical, 1 kHz, 1 VPP, 100% symmetry)
Symmetry 0% to 100%
Pulse
Pulse Width ≥16 ns (limited by the current frequency setting)
Duty Cycle 0.001% to 99.999% (limited by the current frequency setting)
Rising/Falling Edge ≥10 ns (limited by the current frequency setting and pulse width setting)
Overshoot
Typical (1 Vpp)
≤5%
Jitter (rms)
Typical (1 Vpp)
≤5 MHz 2 ppm + 200 ps
> 5 MHz 200 ps
Arbitrary Waveform
Waveform Length 8pts to 8Mpts (16Mpts optional)
Vertical Resolution 14 bits
Specications
All the specications can be guaranteed if the following two conditions are met unless where noted.
· The generator is within the calibration period and has performed self-calibration.
· The generator has been working continuously for at least 30 minutes under the specied temperature (18
℃
~ 28
℃
).
All the specications are guaranteed unless those marked with “typical”.