3. Manual Trigger
A sweep will be generated from the corresponding channel once you press
Trigger when manual trigger is selected.
Note: the menu key Trigger is valid when the Manual Trigger” is selected and
the output of the corresponding channel is enabled.
Press TrigOut to select Leadingor Trailing and the [CH1/Sync/Ext
Mod/Trig/FSK] connector at the rear panel will output trigger signal with
specified edge.
KEY POINTS:
1. When Intor Manualtrigger is selected, press Sweep Trigger
TrigOut to select Off, Leadingor Trailing.
1) When Off is selected, the [CH1/Sync/Ext Mod/Trig/FSK]
connector
at the rear panel will ouput the sync signal which changes from
low level to high level at the start of each sweep and returns to low level
at center frequency point or at the specified mark frequency point.
2) When Leading is selected, the [CH1/Sync/Ext Mod/Trig/FSK]
connector at the rear panel will ouput the trigger signal which is the same
as the sync signal. The trigger signal changes from low level to high level
at the start of each sweep (that is, enable the trigger at the leading edge)
and returns to low level at center frequency point or at the specified mark
frequency point.
3) When Trailing is selected, the [CH1/Sync/Ext Mod/Trig/FSK]
connector at the rear panel will ouput the trigger signal which changes
from high level to low level at the start of each sweep (that is, enable the
trigger at the trailing edge) and returns to high level at center frequency
point or at the specified mark frequency point.
[CH1/Sync/Ext Mod/Trig/FSK]