Quick Start Guide

RIGOL
DG1000Z Quick Guide 5
4. Return to the Previous Menu
Quit the current menu and return to the previous menu.
5. CH1 Output Connector
BNC connector. The nominal output impedance is 50Ω.
When the Output1 is enabled (the backlight goes on), this connector outputs
the waveform based on the current settings of CH1.
6. CH2 Output Connector
BNC connector. The nominal output impedance is 50Ω.
When the Output2 is enabled (the backlight goes on), this connector outputs
the waveform based on the current settings of CH2.
7. Channel Control Area
Used to control the output of CH1.
Press this key to open the output of CH1, the backlight goes
on and the [CH1] connector outputs the waveform based on
the current settings of CH1.
Press this key again to close the output of CH1 and the
backlight goes off.
Used to control the output of CH2.
Press this key to open the output of CH2, the backlight goes
on and the [CH2] connector outputs the waveform based on
the current settings of CH2.
Press this key again to close the output of CH2 and the
backlight goes off.
Used to switch the current selected channel between CH1 and
CH2.
CAUTION
Overvoltage protection of CH1 and CH2 will take effect once any of the
following conditions is met. When overvoltage protection takes effect, a
message will be displayed and the output is disabled.
The input voltage is higher than ±11.5×(1±5%)V (<10kHz) when
the amplitude of the generator is greater than 2Vpp or the DC offset
is greater than |2V
DC
|.
The input voltage is higher than ±3.5×(1±5%)V (<10kHz) when
the amplitude of the generator is lower than or equal to 2Vpp or the
DC offset is lower than or equal to |2V
DC
|.
8. Input Connector for the Signal Measured by Counter
BNC connector. The input impedance is 1MΩ. This connector is used to accept
the signal measured by the counter.