Chapter 2 Front Panel Operations RIGOL
DG1000Z Users Guide 2-59
Sweep Type
DG1000Z provides Linear, Log and Step sweep types and the default is Linear
sweep.
Linear Sweep
In linear sweep type, the output frequency of the instrument varies linearly in the
way of several Hertz per second. The variation is controlled by Start Frequency,
Stop Frequencyand Sweep Time.
When Sweep is enabled, press Type to select Linear. A straight line is displayed
on the waveform on the screen, indicating that the output frequency varies linearly.
Figure 2-7 Linear Sweep
Log Sweep
In Log Sweep type, the output frequency of the instrument varies in a logarithmic
fashion, that is, the output frequency changes in the way of octave per secondor
decade per second. The variation is controlled by Start Frequency, Stop
Frequencyand Sweep Time.
When Log Sweep is enabled, users could set the following parameter:
Start Frequency (
start
F
), Stop Frequency (
stop
F
) and Sweep Time (
sweep
T
).
The function prototype of Log Sweep:
T
current
PF =
current
F
is the instantaneous frequency of the current output.
P
and
T
could be
expressed as shown below by the above-mentioned parameters:
sweepstartstop
TFF
P
/)/lg(
10=
)lg(/)lg( PFtT
start
+=
Wherein,
t
is the time from the start of the sweep and its range is from 0 to
sweep
T
.
When Sweep is enabled, press Type to select Log. An exponential function curve
is displayed on the waveform on the screen, indicating that the output frequency
changes in a logarithmic mode.