Datasheet
Model DG1022Z DG1032Z DG1062Z
Channel 2 2 2
Max Frequency 25 MHz
30 MHz 60 MHz
Sample Rate 200 MSa/s
Waveform
Basic Waveform Sine, Square, Ramp, Pulse, Noise
Built-in Arbitrary Waveform
160 kinds, including Sinc, Exponential Rise, Exponential Fall, ECG, Gauss, HaverSine, Lorentz,
Dual-Tone, etc.
Frequency Characteristics
Sine 1 μHz to 25 MHz 1 μHz to 30 MHz 1 μHz to 60 MHz
Square 1 μHz to 25 MHz 1 μHz to 15 MHz 1 μHz to 25 MHz
Ramp 1 μHz to 500 kHz
1 μHz to 500 kHz
1 μHz to 1 MHz
Pulse 1 μHz to 15 MHz 1 μHz to 15 MHz 1 μHz to 25 MHz
Harmonic 1uHz to 10 MHz 1 μHz to 10 MHz 1uHz to 20 MHz
Noise (-3dB) 25 MHz bandwidth 30 MHz bandwidth 60 MHz bandwidth
Arbitrary Waveform 1 μHz to 10 MHz 1 μHz to 10 MHz 1 μHz to 20 MHz
Resolution 1 μHz
Accuracy ±1 ppm of the setting value, 18
℃
to 28
℃
Sine Wave Spectrum Purity
Harmonic Distortion
Typical (0 dBm)
DC-10 MHz (included): <-65 dBc
10 MHz to 30 MHz (included): <-55 dBc
30 MHz to 60 MHz (included): <-50 dBc
Total Harmonic Distortion <0.075% (10 Hz to 20 kHz, 0 dBm)
Spurious (non-harmonic)
Typical (0 dBm)
≤10 MHz: <-70 dBc
>10 MHz: <-70 dBc + 6 dB/octave
Phase Noise
Typical (0 dBm, 10 kHz offset)
10 MHz: <-125 dBc/Hz
Signal Characteristics
Square
Rise/Fall Time
Typical (1 Vpp)
<10ns
Overshoot
Typical (100 kHz, 1 Vpp)
≤5%
Duty Cycle 0.01% to 99.99% (limited by the current frequency setting)
Non-symmetry 1% of the period + 5 ns
Jitter (rms)
Typical (1 Vpp)
≤5 MHz: 2 ppm + 200 ps
> 5 MHz: 200 ps
Ramp
Linearity ≤1% of peak output (typical, 1 kHz, 1 VPP, 100% symmetry)
Symmetry 0% to 100%
Pulse
Pulse Width 16ns to 999.999 982 118ks (limited by the current frequency setting)
Duty Cycle 0.001% to 99.999% (limited by the current frequency setting)
Rising/Falling Edge ≥10 ns (limited by the current frequency setting and pulse width setting)
Overshoot
Typical (1 Vpp)
≤5%
Jitter (rms)
Typical (1 Vpp)
≤5 MHz 2 ppm + 200 ps
> 5 MHz 200 ps
Arbitrary Waveform
Waveform Length
2Mpts (std.)
16Mpts (opt.)
8Mpts (std.)
16Mpts (opt.)
8Mpts (std.)
16Mpts (opt.)
Specications
All the specications can be guaranteed if the following two conditions are met unless where noted.
· The generator is within the calibration period and has performed self-calibration.
· The generator has been working continuously for at least 30 minutes under the specied temperature (18
℃
~ 28
℃
).
All the specications are guaranteed unless those marked with “typical”.
3
RIGOL