User manual
DG1022 Commands System RIGOL
Programming Guide for DG1022
2-7
50% automaticly.
Explanations
If the parameters you set are less than three, the sequence
would be: <frequency>, <amplitude>, <offset>.
The default units of <frequency>, <amplitude>, <offset> are:
Hz, Vpp, V
DC
.
Example
APPL:SQU:CH2 1000,5.0,-1
11. APPLy:RAMP:CH2
Command
Format
APPLy:RAMP:CH2 [<frequency>[,<amplitude>[,<offset>]]]
Function
Generate a ramp wave with specific frequency, amplitude and DC
offset via CH2 and cover the current symmetry settings and select
50% automaticly.
Explanations
If the parameters you set are less than three, the sequence
would be: <frequency>, <amplitude>, <offset>.
The default units of <frequency>, <amplitude>, <offset> are:
Hz, Vpp, V
DC
.
Example
APPL:RAMP:CH2 1000,5.0,0.5
12. APPLy:PULSe:CH2
Command
Format
APPLy:PULSe:CH2 [<frequency>[,<amplitude>[,<offset>]]]
Function
Generate a pulse wave with specific frequency, amplitude and DC
offset via CH2.
Explanations
If the parameters you set are less than three, the sequence
would be: <frequency>, <amplitude>, <offset>.
The default units of <frequency>, <amplitude>, <offset> are:
Hz, Vpp, V
DC
.
Example
APPL:PULS:CH2 1000,5.0,0.5
13. APPLy:NOISe:CH2
Command
Format
APPLy:NOISe:CH2[<frequency|DEFault>[,<amplitude>[,<offset>]]]
Function
Generate Gaussian noise with specific amplitude and DC offset via
CH2.
Explanations
Although the frequency parameter made no impression on this
command, a value or “DEFault” must be specified. (noise