Users Manual

85
Calibration (Optional)
If you are experiencing rapid glucose changes
(more than 2 mg/dL per minute), the sensor
readings displayed may be less accurate and not
as timely.
When you see the symbol on your receiver
(HR310) display, you must check your BG value
with a BG meter before making any treatment
decisions. Sensor readings may be less accurate
and may not reflect your current glucose levels.
CALIBRATE
When you calibrate, take a fingerstick measurement from
your BG meter then enter the value according to the
following steps:
1. From the Calibrate screen, tap [Calibrate Now].
1