Getting to Know Your RIGHTEST CGMs
SAFETY FEATURES
The RIGHTEST CGMs offers a number of important safety features when you use it.
These features include:
Receiver (HR310)’s alarm/alert includes visual notification, vibrations and sound, depending on
your personalized settings.
When you are out of your target glucose range, the receiver (HR310) alerts you.
Receiver (HR310) warns you if your glucose level falls below 54 mg/dL; or when your sensor
reading is below 70 mg/dL and continuously falling for the following 2 readings (or 10 minutes).
Urgent alert/alarm settings at 54 mg/dL or below; or below 70 mg/dL and continuously falling for
the following 2 readings (or 10 minutes) cannot be changed or turned off.
Receiver (HR310) notifies you when a sensor has failed, expired or when there are system errors.