Users Manual
21
Getting to Know Your RIGHTEST CGMs
SAFETY FEATURES
The RIGHTEST CGMs offers a number of important safety features when you use it.
These features include:
• Receiver (HR310)’s alarm/alert includes visual notification, vibrations and sound, depending on
your personalized settings.
• When you are out of your target glucose range, the receiver (HR310) alerts you.
• Receiver (HR310) warns you if your glucose level falls below 54 mg/dL; or when your sensor
reading is below 70 mg/dL and continuously falling for the following 2 readings (or 10 minutes).
• Urgent alert/alarm settings at 54 mg/dL or below; or below 70 mg/dL and continuously falling for
the following 2 readings (or 10 minutes) cannot be changed or turned off.
• Receiver (HR310) notifies you when a sensor has failed, expired or when there are system errors.