➔ Rudolf Riester GmbH Postfach 35 • DE-72417 Jungingen Germany Tel.: +49 (0)74 77-92 70-0 Fax: +49 (0)74 77-92 70-70 info@riester.de www.riester.de Operator´s manual Gebrauchsanweisung Mode d' emploi Instrucciones para el uso Istruzioni per l'uso Инструкции по зксплуатации 99151 Rev.
ENGLISH OPERATOR’S MANUAL Pulse Oximeter ri-fox N
General Description Oxygen Saturation is a very important parameter for the Respiratory Circulation System. Many respiratory diseases can result in oxygen saturation being lowered in human blood. Additionally, the following factors can reduce oxygen saturation: organ dysfunction caused by anesthesia, intensive postoperative trauma, injuries caused by some medical examinations. That situation might result in light-headedness, asthenia, and vomiting.
Product Operation Scope ri-fox N is a portable non-invasive, spot-check, oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients at home and hospital (including clinical use in internal medicine/surgery, anesthesia, etc). The device is not intended for continuous monitoring. The ri-fox N requires no routine calibration or maintenance other than replacement of batteries. Technical Specifications 1. Display: LED 2.
4. Power Requirements: Two AAA alkaline Batteries Power consumption: Less than 30mA Low power indication: Battery Life: Two AAA 1.5V, 800mAh alkaline batteries could be continuously operated as long as 30 hours. 5. Dimension: Length: 66mm Width: 39mm Height: 32mm Weight: 34g (without batteries) 6. Environment Requirements: Operation Temperature: 5°C~40°C Storage Temperature: -20~55°C Ambient Humidity: ≤80%, no condensation in operation; ≤93%, no condensation in storage 7.
Precautions for use 1. Do not use the pulse oximeter in an MRI or CT environment 2. Do not use the pulse oximeter in situations where alarms are required. The device has no alarms. 3. Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere. 4. The pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms. 5.
9. Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or immersing the sensors in liquid may cause inaccurate readings. 10. SpO2 measurements may be adversely affected in the presence of high ambient light. Shield the sensor area if necessary. 11. Excessive patient movement may cause inaccurate readings. 12. Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line may cause inaccurate readings. 13.
Product specifications 1. 2. 3. 4. 5. Easy to use. Small in volume, light in weight. Low power consumption. Low voltage warning will be indicated in visual window when battery voltage is low. The product will automatically be powered off when no signal is registered for longer than 8 seconds. Operation Instructions 1. 2. 3. 4. 5. 6. Install two AAA batteries into the battery cassette, and then close its cover. Open the clamp as shown in the picture below.
Brief Description of Front Panel Pulse low indicator Oxygen Saturation %SpO2 Pulse Bar graph Pulse Rate PR BPM The Pulse Bar graph displays the strength of the pulse rate signal. The height of the bar graph shows the patient’s pulse strength. Battery Installation Press the button down! 1. Put the two AAA batteries into battery cassette in correct polarities. 2.
Maintenance and Storage 1. Replace batteries in time when low voltage lamp is lighted. 2. Clean surface of the ri-fox N before it is used in diagnosis for patients. 3. Remove batteries inside the battery cassette if the oximeter will not be used for a long time. 4. It is best to preserve the product in a place where ambient temperature is –20~55°C and humidity is <93%, no condensation. 5. It is recommended that the product should be kept in a dry environment anytime.
Declaration EMC of this product complies with IEC60601-1-2 standard. Portable and mobile high frequency appliances can influence medical electronic devices. Medical electronic devices need special safety requirements concerning EMC. You can get a detailed description of the EMC-requirements at www.riester.de. The materials which the user can come into contact have no toxicity and no action on tissues, comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Possible Problems and resolutions Problems Possible reason SpO2 or PR 1.Finger is not plugged correctly can not be 2.Patient’s Oxyhemoglobin value is shown normally too low to be measured. 1.Finger might not be plugged SpO2 or PR is deeply enough. shown unstably 2.Finger is trembling or patient’s body is in movement status. 1.Power of batteries might be The Oximeter inadequate or not be there at all. can not be 2.Batteries might be installed powered on incorrectly. 3.The Oximeter might be damaged. 1.
Symbol Definitions Symbol Definition Type BF applied part Attention, consult accompanying documents. ﹪SpO2 PR BPM Oxygen saturation Pulse rate (BPM) Low power indication NOT for continuous monitoring SN Serial No. Note: The illustrations used in this manual may differ slightly from the appearance of the actual product.
Disinfection Please use medical alcohol to clean the plastic material touching the finger inside of the pulse oximeter, and clean the test sensor using alcohol before and after operation. Warranty This product has been manufactured under the strictest quality standards and has undergone a thorough final quality check before leaving our factory.