Specifications
70  RL-TA Series Operation Manual 
18  Use Printers 
18.1 Setting up the Printer 
(1)  Use a dot matrix or thermal RS-232 printer with the balance. 
(2)  Set proper print functions (print control) with the balance referring to the instruction manual for your 
printer. 
(3)  Make the baud rate and other settings compatible between the balance and the printer. 
18.2 Outputting calibration results 
18.3 Output of Measurement Data in Compliance with ISO/GLP/GMP 
1.  Set Function 1. 
Set the ISO/GLP/GMP-compatible item to [F GLP 1] and the GLP-compatible item for 
measurement data to [F2. od. 1]. 
2.  Header output. 
Press and hold the Print key for a few seconds. 
[HEAd] is displayed and the header is then output. 
3.  Measurement data output. 
Measurement data can be output at any time during measurements. 
4.  Footer output. 
Press and hold  the  Print  key for a few seconds  after the 
measurement is completed. 
The display indicates [Foot] and a footer is output. 
(Continued on next page.) 