Specifications
    RL-TA Series Operation Manual  67 
17.4.8  Setting limit values 
Command Main Body 
Description  Value  Response 
first 
character 
second 
character 
ASCII code 
L  A  4CH  41H 
First setting 
point/lower limit 
Limit 
values 
A00: Successful 
completion 
E01:Command error 
E02:Set value error 
L  B  4CH  42H 
Second setting 
point/upper limit 
L  C  4CH  43H  Reference value 
L  D  4CH  44H  Third setting point 
L  E  4CH  45H  Forth setting point 
1.  Insert a comma between command entries and limit values; e.g., LA, nnn. 
2.  Enter numerical values only.   You do not need to enter units. 
3.  In weighing machine mode, the unit set in Unit A is applied to the entered value.   For example, when 
“LA, 20.00” is transmitted, it is recognized as 20.00 g if gram is selected in Unit A, and 20.00 ct if 
carat is selected in Unit A. 
4.  The units used for parts counting, percentage weighing, and unit converting are PCS, %, and # 
respectively. 
17.4.9  Sample input commands 
T(SP)(CR)(LF)  Set tare range or adjust to zero-point. 
O1(CR)(LF)  Set the balance to continuous output. 
O8(CR)(LF)  Output data (once immediately). 
IA,01,30,00(CR)(LF)  Set an interval of 01 hrs: 30 min: 00 sec. 
OA(CR)(LF)  Start the interval function. 
DD(CR)(LF)  Output the date. 
DT(CR)(LF)  Output the time. 
LA,80.5(CR)(LF)  Set the first setting point to 80.5.   (The unit varies depending on the weighing 
mode.) 










