Specifications

RL-TA Series Operation Manual 67
17.4.8 Setting limit values
Command Main Body
Description Value Response
first
character
second
character
ASCII code
L A 4CH 41H
First setting
point/lower limit
Limit
values
A00: Successful
completion
E01:Command error
E02:Set value error
L B 4CH 42H
Second setting
point/upper limit
L C 4CH 43H Reference value
L D 4CH 44H Third setting point
L E 4CH 45H Forth setting point
1. Insert a comma between command entries and limit values; e.g., LA, nnn.
2. Enter numerical values only. You do not need to enter units.
3. In weighing machine mode, the unit set in Unit A is applied to the entered value. For example, when
“LA, 20.00” is transmitted, it is recognized as 20.00 g if gram is selected in Unit A, and 20.00 ct if
carat is selected in Unit A.
4. The units used for parts counting, percentage weighing, and unit converting are PCS, %, and #
respectively.
17.4.9 Sample input commands
T(SP)(CR)(LF) Set tare range or adjust to zero-point.
O1(CR)(LF) Set the balance to continuous output.
O8(CR)(LF) Output data (once immediately).
IA,01,30,00(CR)(LF) Set an interval of 01 hrs: 30 min: 00 sec.
OA(CR)(LF) Start the interval function.
DD(CR)(LF) Output the date.
DT(CR)(LF) Output the time.
LA,80.5(CR)(LF) Set the first setting point to 80.5. (The unit varies depending on the weighing
mode.)