Specifications
RL-TA Series Operation Manual 61
17.3.5 Judgment result when the limit function is enabled (S1: 1 character)
S1
ASCII Code
Description
L
4CH
Too little (LO)
1- or 2-point scale G 47H Proper (OK)
H 48H Too much (HI)
1 31H Rank 1
3- or 4-point scale
2 32H Rank 2
3 33H Rank 3
4
34H
Rank 4
5 35H Rank 5
T 54H Cumulative value
Data type
U 55H Unit weight
(SP) 20H No judgment result or no data type specified
d
64H
Gross
17.3.6 Status (S2:1 character)
S2 ASCII Code Description
S 53H Data stable *
U 55H Data unstable *
E 45H
Data error (Indicates that data other than S2 is invalid.)
([o - Err], [u - Err])
(SP) 20H No status specified
* Data that is independent of whether the weighing condition is stable or not, such as cumulative values
and unit weights, is independent of whether S2 is S or U when it is output.
17.3.7 Output data other than measurement data
The following data is accompanied by our proprietary printer control commands prior to and after the data
when output.
(1) Interval output
When each interval output is started or stopped, a header and a footer are output.
Header
‘
-’ is output for 15 characters and terminators (CR=0DH, LF=0AH).
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
- - - - - - - - - - - - - - -
Footer
Two linefeeds are inserted.
(Continued on next page.)