Specifications
44  RL-TA Series Operation Manual 
3.  When [
on F.S.
] is displayed, put a weight on the pan. 
The balance starts adjusting the weighing capacity.  
The display indicates [buSY]→[End] in sequence. When span 
adjustment is completed, the balance goes back to the normal display. 
14.3 Span Test with Built-In Weights 
1.  Press the Cal key. 
[t. int] is displayed. 
2.  When [
PuSH C
] is displayed, press the Cal key again. 
The built-in weight unit starts to operate, and the balance will automatically start span test. 
The display changes in the order of [t.Int]→[t.0]→[  t.F.5]→[dIFF]→[Weight error]. 
Pressing any key returns the display to measurement mode. 
14.4 Span Test with External Weight 
1.  Press the Cal key. 
[t. Eht] is displayed. 
2.  When [
PuSH C
] is displayed, press the Cal key again. 
The display first indicates [t. EHt]→[on 0], and the balance starts testing the zero-point. 
 (If a weight error is used, the corresponding weight error entered will be displayed for a few 
seconds.) 
3.  When [
on F.S.
] is displayed, put a weight on the pan. 
 The balance starts a weighing capacity point test. 
The display first indicates [dIFF] and then an error. 
When you press a key, the balance goes back into the measurement 
mode. 
* After [dIFF] is completed, the definition of a weight error is as follows: 
 Weight error = True value – Current weight 
In other words, when a weight error is more than “0”, the displayed weight 
is less than the actual weight. 