Specifications
    RL-TA Series Operation Manual  35 
When 3- or 4-points are set, triangles [] are displayed in four or five levels according to the judgment 
result.   
Judgment Result  (3- or 4-point setting) 
Rank 5 
(4-point setting) 
Fourth point ≤ Weight 
Rank 4 
Third point ≤ Weight < Fourth point 
Rank 3 
Second point ≤ Weight < Third point 
Rank 2 
First point ≤ Weight < Second point 
Rank 1  Weight < First point 
The [<] sign stays on in the range where the [] sign is displayed depending on the number of points set. 
13.4 Judge by Absolute Values 
13.4.1  Set two limit values by putting actual samples on the balance − Judge by Absolute Values − 
1.  Start the limit function. 
Press and hold the Target key for a few seconds.    
When the display indicates [L. SEt], release the   
key. 
The currently saved lower limit is displayed. 
2.  Put the lower limit sample on the balance. 
Put the lower limit sample on the pan. 
3.  Save the lower limit. 
Press the Function key. 
When a lower limit is saved, the value is displayed  
for a while. 
(When one point is set, the setting is completed.) 
4.  Set an upper limit. 
The display then indicates [H. SEt], and you can now set an upper limit. 
The currently saved upper limit is displayed. 
5.  Put the upper limit sample on the balance. 
Put the upper limit sample on the pan. 
6.  Save the upper limit. 
Press the Function key. 
(Continued on next page.) 
n