10  Unit Converting 
The function to convert units multiplies the weight of a sample on the balance by the saved coefficient, and 
displays the result. 
* Select [1    Set . 4] (Unit Converting) in Function 1. 
1.  Displaying the coefficient setting. 
Make sure that the [#] sign is displayed and press and hold the 
Function key for a few seconds. 
(This operation may be performed with a container placed on 
the balance.) 
The display then blinks [C. SEt]. 
The previously-saved reference sample weight blinks. 
2.  Enter a coefficient. 
Press the Zero/Tare key. 
The digit furthest to the right blinks. 
3.  Changing numbers. 
Pressing  the  Zero/Tare  key  changes  the number as 
shown on the right. 
4.  Select digit places to specify a coefficient. 
Pressing the Function key shifts the blinking digit to the left, 
allowing you to set a higher-order digit. 
After a number is set, the minus sign [M] blinks. Pressing the 
key again makes the digit furthest to the right blink. 
5.  Save the coefficient. 
Press the Target key.* 
(* Pressing the Print key cancels the setting  
of a reference sample weight.) 
6.  Put a sample to weigh on the balance. 
Put a sample to weigh on the balance. 
The weight of the sample is multiplied by the coefficient, and 
the result is displayed. 