Specifications
    RL-TA Series Operation Manual  21 
8  Parts Counting 
When a specified number of samples are put on the balance, the balance divides the total weight of the samples 
by the number of samples to obtain the average sample weight. The balance stores the average sample weight 
based on the number of samples initially set. Additional samples, up to three times the initial number, can be 
added and the balance will automatically update the average sample weight. This mechanism allows accurate 
counting. 
* Select
 [1. Set 2] (Parts Counting) in Function 1. 
1.  Start sampling. 
Make sure that the [Pcs] sign is displayed and press and hold 
the Function key for a few seconds. 
(This operation may be performed with a container placed on 
the balance.) 
When the display indicates [U. SEt], 
release the key. 
The display then blinks [on 10]. 
This indicates to put 10 samples on the balance. 
2.  Change the number of samples as required. 
The number of samples can be changed by pressing the 
Zero/Tare key. 
If the samples vary considerably in size or are lightweight, or if 
more accurate counting is required, press the Zero/Tare key to 
sufficiently increase the number of samples. 
If you do not need to change the number, skip this step. 
3.  Put a sample on the balance. 
Put as many samples as displayed in [on **] on the balance 
and press the Function key. 
The display then blinks [10
 Pcs
]. 
4.  Put additional samples on the balance. 
Add  up to  three times the displayed number of  samples and 
wait until the display stabilizes  again. The [+]  sign is 
displayed and the memory is updated. 
Every time the [+] sign is displayed, you may add more samples. 
(Continued on next page.) 