Specifications
76  RL-TS Series Operation Manual 
-  When either an O8 or O9 command is executed, it returns to the state of “O0”. 
(3)  Set measurement mode command 
C1  C2  Code (C1)  Code (C2)  Description  Response 
M  1  4DH  31H  Set to Mode 1 
A00: Successful completion 
E01:  Command error 
E02:  Error 
M  2  4DH  32H  Set to Mode 2 
M  3  4DH  33H  Set to Mode 3 
M  4  4DH  34H  Set to Mode 4 
*  The  measurement mode to be activated by the above mode settings 1 to 4 depends on the weighing mode 
currently in use. 
- Relationship between weighing mode and mode setting 
Mode 
Weighing 
machine 
Parts 
counting 
Percentage 
weighing 
Unit 
converting 
Gravimeter 
Animal 
weighing 
Mode 1 
Weight 
measuring 
Weight 
measuring 
Weight 
measuring 
Weight 
measuring 
Error  Error 
Mode 2  Gross weight 
Parts 
counting 
Percentage 
measuring 
Coefficient 
multiplying 
Error  Error 
Mode 3 
Cumulative 
weight*1 
Cumulative 
count*1 
Cumulative 
percent*1 
Cumulative 
sum*1 
Error  Error 
Mode 4 
Display in 
unit B*2 
Average unit 
weight 
Error  Error  Error  Error 
*1: Mode 3 (M3) can be specified only when the addition function is to be used. If the addition function is not 
enabled, it results in an error. 
*2: If a unit is NOT specified for unit B, the balance is set for the weight measuring mode. When you specify a 
mode that is not supported by the current Weighing Mode, an error is returned from the balance. 
(4) Span adjustment/test command 
C1  C2  Code (C1)  Code (C2)  Description  Response 
C  0  43H  30H 
Disables command 
inputs.*1 
A00: Successful completion 
E01:  Command error 
E02:  Operation is disabled. 
E03:  Cancelled by operation 
E04:  Abnormal completion 
C  3  43H  33H 
Span adjustment with 
external weight 
C  4  43H  34H 
Span test with external 
weight 
*1: A span adjustment and test commands will also be disabled. 
*  This command takes time because the balance sends back a response after an appropriate operation is 
completed. 
*  If set to “[Cal] key disabled” (0) in the function setting “7.CA.,” a span adjustment and test command does 
not work. 
66BResponse 
Upon receiving an input command, the balance sends out a response. 