Specifications
    RL-TS Series Operation Manual  75 
  Command format 
(1)  Tare range (zero-point adjustment) command 
C1  C2  Code (C1)  Code (C2)  Description  Value   Response 
T  (SP)  54H  20H 
-  Tare range setting 
-  Zero-point 
adjustment 
None 
A00: Successful completion 
E01:  Command error 
E04:  A tare range (zero-point 
adjustment) cannot be set 
(Range violation, weight error, 
etc.). 
(2)  Set output control command 
C1  C2  Code (C1)  Code (C2)  Description 
Response 
O  0  4FH  -{}-30H  Stop output. 
A00: Successful 
completion 
E01:  Command 
error 
O  1  4FH  31H  Output continuously at all times. 
O  2  4FH  32H 
Output continuously if stable (Stop output if 
unstable). 
O  3  4FH  33H 
Output once when the [Print] key is pressed 
(whether the balance is stable or unstable). 
O  4  4FH  34H 
Output once when the balance is stable. Output 
when a sample is lifted to cause the display to 
indicate a value below zero, and then another 
sample is placed to make the balance stable. 
O  5  4FH  35H 
Output once when the balance is stable. Stop 
output when unstable. Output once when the 
balance is stabilized again (the output includes 
zero) even if it is not reloaded. 
O  6  4FH  36H 
Output once when the balance is stable. Output 
continuously when unstable. Output is stopped 
after a single output when the balance is stable 
even if it is not reloaded. 
O  7  4FH  37H 
Output once when the [Print] key is pressed if 
the balance is stable. 
O  8  4FH  38H  Output once immediately. 
O  9  4FH  39H  Output once after stabilized. 
-  Commands O0 to O7 have the same workings as the output control set by the function setting. 
-  Commands O8 and O9 are used to request data from the balance. 
-  Once executed, O0 to O7 commands are held. However, the status is reset to the function setting when the 
balance is turned on again. 