5.2 Minimum Readability Setting 
Use this function to set the minimum readability. The larger the minimum readability becomes, the less the 
balance is affected by external influences. In addition, it takes less time for the balance reading to become 
stable. 
Each unit has different minimum readability. 
1 
Set to the function setting mode. 
(Refer to “Section 2-5: Function Setting 
Basics (P. 
X22X).) 
Press and hold the [Function] key. After “
Func
” is 
displayed, release the finger. 
2 
Select the minimum readability. 
Press the [Function] key several times to select 
“b2.d.A.” 
Select from 1 through 5 by pressing the [Zero/Tare] key. 
3 
Save the setting. 
Press the [Target] key. 
The balance goes back to normal display. 
To set the minimum readability of unit B, select “b4.d.b” in step 2. 
You can also set the same unit for unit A and unit B, and set different minimum readability, so that unit 
A and unit B can be used to switch the minimum readability. 
  Minimum readability example 
Setting 
value 
TS-220~820  TS-2200~8200  TS-15K 
  g  kg  ct  mom  g  kg  ct  mom  g  kg  ct  mom
1  0.01  0.00001  0.05
0.005  0.1  0.0001  0.5  0.05  1  0.001  5  0.5 
2  0.02  0.00002  0.1  0.01  0.2  0.0002  1  0.1  2  0.002  10  1 
3  0.05  0.00005  0.2  0.02  0.5  0.0005  2  0.2  5  0.005  10  2 
4  0.1  0.0001  0.5  0.05  1  0.001  5  0.5  10  0.01  10  5 
5  0.2  0.0002  1  0.1  2  0.002  10  1  10  0.02  10  10 