Specifications

RL-TS Series Operation Manual 53
22B4.2 Testing the Balance
The amount of offset with respect to the reference weight can be checked. This is called “span test.”
Performing a span test will not calibrate the balance.
Use a weight for calibration that weighs 50% of the weighing capac
ity or heavier. To calibrate more
accurately, use a weight that is equivalent to the weighing capacity.
Please contact us if you wish inquire about or place an order for calibration weights.
1
Set to the function setting mode.
(Refer to “Section 2-5: Function Setting
Basics (P.
X22X).)
Press and hold the [Function] key. After “Func” is
displayed, release the finger.
2
Set to the span test mode.
Press the [Function] key several times to select “7.CA.
Press the [Zero/Tare] key several times to select “7.CA 4.”
3
Save the setting.
Press the [Target] key.
The balance goes back to normal display.
4
Start span test.
<Press and hold>
Keep holding down the [Function] key untiltESt” is
displayed.
After zero-point testing (on 0” blinks), the display changes
to “on F.S.
5
Place the weight on the pan.
After “on F.S.” is displayed, place the weight on the pan.
The test starts.
6
The error is displayed.
The display first indicates “dIFF and then the weight error
of the balance.
A weight error is as follows:
Weight error = True value Current weight
When a weight error is more than “0”, the displayed weight is
less than the actual weight.
Press any key to return to normal display.
Reference