Specifications
    RL-TS Series Operation Manual  53 
22B4.2 Testing the Balance 
The amount of offset with respect to the reference weight can be checked. This is called “span test.” 
Performing a span test will not calibrate the balance. 
Use a weight for calibration that weighs 50% of the weighing capac
ity or heavier. To calibrate more 
accurately, use a weight that is equivalent to the weighing capacity. 
Please contact us if you wish inquire about or place an order for calibration weights. 
1 
Set to the function setting mode. 
(Refer to “Section 2-5: Function Setting 
Basics (P. 
X22X).) 
Press and hold the [Function] key. After “Func” is 
displayed, release the finger. 
2 
Set to the span test mode. 
Press the [Function] key several times to select “7.CA.” 
Press the [Zero/Tare] key several times to select “7.CA 4.” 
3 
Save the setting. 
Press the [Target] key. 
The balance goes back to normal display. 
4 
Start span test. 
<Press and hold> 
Keep holding down the [Function] key until “tESt” is 
displayed. 
After zero-point testing (“on 0” blinks), the display changes 
to “on F.S.” 
5 
Place the weight on the pan. 
After “on F.S.” is displayed, place the weight on the pan. 
The test starts. 
6 
The error is displayed. 
The display first indicates “dIFF” and then the weight error 
of the balance. 
A weight error is as follows: 
Weight error = True value – Current weight 
When a weight error is more than “0”, the displayed weight is 
less than the actual weight. 
Press any key to return to normal display. 