Specifications
44  RL-TS Series Operation Manual 
53BLimit function setting 
First, set the limit function. Then, set limit values. 
1 
Set to the function setting mode.
(Refer to “Section 2-5: Function Setting Basics 
(P. 
X22X).) 
Press and hold the [Function] key. After 
“Func” is displayed, release the finger. 
2 
Select Limit Function. 
Press the [Function] key several times to select 
“2.SEL.” 
Press the [Zero/Tare] key to select “2.SEL 
2.” 
When using both the cumulate and limit 
functions together, select “2.SEt 3..” 
3 
Set the judgment condition. 
Press the [Function] key several times to select 
“21.Co.” 
Press the [Zero/Tare] key to select the judgment 
condition. 
1: Always judge (even when the balance is 
unstable). 
2: Judge only when the balance is stable. 
4 
Set the judgment range to cover. 
Press the [Function] key several times to select 
“22.Li.” 
Press the [Zero/Tare] key to select the judgment 
range to cover. 
0:  Detect when the limit is exceeded by more 
than five divisions. 
1:  Detect both when the limit is exceeded 
and when it is not reached. 
5 
Set the point scale to be set. 
Press the [Function] key several times to select 
“23.Pi .” 
Press the [Zero/Tare] key to select the point 
scale to be set. 
1:  1-point scale (OK and LO are judged.) 
2:  Upper and lower limits are set (HI, OK 
and LO are judged). 
6 
Set the judgment method. 
Press the [Function] key several times to select 
“24.tP.” 
Press the [Zero/Tare] key to select the judgment 
method. 
1:  Judge by absolute values. 
2:  Judge by deviation values. 