Specifications
38  RL-TS Series Operation Manual 
18B3.6 Weighing an Animal 
The balance can accurately weigh animals and other samples that move during measurement. 
Even if animals and other such moving samples move during measurement, if weight variations fit within a 
set value range, the measured value is held (fixed) and displayed. 
A weight variation range for value holding can be set with “16.Ud.
”
By changing minimum readability, a stability detection range can be also changed 
with a combined use of “16.Ud”  (Refer to “Section 5-2: Minimum Readability 
Setting” (P. 
X56X)). 
1 
Set to the function setting mode.
(Refer to “Section 2-5: Function Setting Basics 
(P. 
X22X).) 
Press and hold the [Function] key. After “Func” is 
displayed, release the finger. 
2 
Select “Animal Weighing.” 
Press the [Zero/Tare] key several times to select  
“1.SEt 6.” 
3 
Save the weighing mode. 
Press the [Target] key. 
The “animal weighing” mode is set, displaying “ .” 
4 
Set the response speed according to the 
animal move. 
Set the response speed using the [Target] key. 
The response speed is indicated with the position to which 
the pointer  located in the right side of the display points. 
5 
Place the sample animal. 
When weight variations fit within the set range, the value is 
held. Holding the readout is reported with “ ” shown in 
the display. 
When automatic tare range setting is ON, a tare range is 
automatically set after the animal is removed and weight 
variations fit within a set range. When automatic tare range 
setting is OFF, a held value is displayed until the [Zero/Tare] 
key is pressed. 
Fast (quiet) 
Middle (in-between) 
Slow (wild) 
Reference