Specifications
22  RL-TS Series Operation Manual 
12B2.5 Function Setting Basics 
To set the settings of the balance, use the function keys. 
This section describes the basic operations of function setting 
Pressing the [Print] key when a function is being set can cancel the setting 
and return to measurement mode. 
1 
Set to the function setting mode. 
After display changes, 
release the finger. 
Hold down the [Function] key and release the 
finger after display has changed to “Func.” 
The first function item is displayed. 
Keeping the [Function] key held down will 
switch the balance to another mode. If this 
happens, press the [Print] key to cancel the 
setting and redo the setting from the beginning. 
2 
Select a setting item. 
By pressing the [Function] key, select an item to 
be set. 
Example: By pressing the [Function] key once, 
select “2.SEL” (additional function). The first 
setting “2.SEL 0” is displayed. 
3 
Select a setting. 
Pressing the [Zero/Tare] key selects a setting. 
Pressing the key toggles the settings in turn. 
After the last setting is displayed, the next 
setting displayed is the first one. 
Example: By pressing the [Zero/Tare] key twice, 
select “2.SEL 2.” 
4 
Save the setting. 
Press the [Target] key to complete the setting. 
The display returns to showing the sample’s 
normal weight. 
To cancel, press the [Print] key. 
Reference