Specifications

12 RL-TC Series Operation Manual
4.2 Description of Functions
Item
Set Value
Description
Bar graph display
1
..b.G..
0
Disable
1
Enable
Limit function
2
.SEL
0
Disable
1
Enable
Displayed only when limit
function is activated
Judgement
condition
21
.Co.
1
Always judge (judges even when the balance is unstable)
2
Judge only when the balance is stable (does not judge
if the balance is unstable)
Judgement
range
22
.L
0
Ranges beyond +5 graduation is judged (ranges +5
graduation or below, including negative ranges,
are not
judged.)
1
The entire range is judged (the entire range, including
the negative, is judged).
Number of
points for
judgement
23.P
1
One-point setup (judges between OK and LO)
2
Upper-limit and lower-limit values are set up (judges
among HI, OK and LO).
Auto-zero
(zero-tracking)
3
.A.O
0
Disable
This function automatically sets the zero point
exactly to zero to prevent slight deviations.
1
Enable
Auto power-off
4
.A.P.
0
Disable (balance operates
continuously)
This function is available
only when the balance is
battery
-operated.
1
Enable (balance powers off in
approximately three minutes)
Response speed
5
.rE.
0
Measurement by consecutive weighings.
1
Fast
Slow
2
3
4
5
Stability parameters
6
. 5.d.
1
Wide (mild)
Narrow (strict)
2
3
4
Interface
7
. 1.F.
0
Disable input/output
1
Six-digit numeric format
2
Seven-digit numeric format
Setup of units of
measurement to be
displayed
Register selected
measuring
units with
Function key.
1 01
[ g ]
2 14
[ ] (ct)
15
[ ] (oz)
16
[ ] (lb)
17
[ ] (ozt)
18
[ ] (dwt)
*
19
[ Lower right ] (grain)
Items marked are the default factory settings.
15: default settings [81.5.u][85.5.u.]
*Not available in TC-12K.