Troubleshooting guide
24 Summit SB-1000 Bariatric Wheelchair Scale Installation Manual
Table 8-2 lists the various display messages when testing the scale.
NOTE: <-> means the display toggles between the two values. 
Periodic Maintenance
The space between the platform side and pit frame, and the surface beneath the platform must be periodically 
cleaned to prevent debris build up. More frequent cleaning of these areas is necessary with scales mounted in 
pits.
Do not attempt to use scales with load cells that are not hermetically sealed in washdown applications. Water 
damage is a common cause of failure in non-hermetically-sealed load cells. Use care with high pressure steam 
washdowns for hermetically-sealed load cells. The steam may not damage the load cells, but the elevated 
temperatures may cause incorrect readings until the unit cools to room temperature.
Step Function Display Available Parameters
1
Turn the scale on. 
While StArt is 
displayed, press and 
hold the 
REWEIGH 
key.
This enters into 
the test mode of 
the scale.
StArt The scale automatically advances to Step 
2.
2
Identifies the 
software ID
IdEnt <-> 11007 Press the REWEIGH key to advance to 
the next step.
3
Shows the current 
weight value
tESt <-> 0.0 To zero the display, press the Kg-Lb key.
To advance to the next step, press the 
REWEIGH key.
4
Shows the current 
A/D count
A-d <-> XXXX To advance to the next step, press the 
REWEIGH key.
5
Checks for current 
battery level
bAt <-> XXX or nO bAt If the nO bAt is displayed, there are not 
batteries in the unit, or the unit is 
operating on its external AC adaptor. 
To advance to the next step, press the 
REWEIGH key.
6
This shows the 
calibration factor.
FACtOr <-> XXXXX Press the REWEIGH key to return to the 
top of the test menu. 
To exit test mode, press the REWEIGH 
and Kg-Lb key simultaneously, or turn off 
the indicator. 