Troubleshooting guide

24 Summit SB-1000 Bariatric Wheelchair Scale Installation Manual
Table 8-2 lists the various display messages when testing the scale.
NOTE: <-> means the display toggles between the two values.
Periodic Maintenance
The space between the platform side and pit frame, and the surface beneath the platform must be periodically
cleaned to prevent debris build up. More frequent cleaning of these areas is necessary with scales mounted in
pits.
Do not attempt to use scales with load cells that are not hermetically sealed in washdown applications. Water
damage is a common cause of failure in non-hermetically-sealed load cells. Use care with high pressure steam
washdowns for hermetically-sealed load cells. The steam may not damage the load cells, but the elevated
temperatures may cause incorrect readings until the unit cools to room temperature.
Step Function Display Available Parameters
1
Turn the scale on.
While StArt is
displayed, press and
hold the
REWEIGH
key.
This enters into
the test mode of
the scale.
StArt The scale automatically advances to Step
2.
2
Identifies the
software ID
IdEnt <-> 11007 Press the REWEIGH key to advance to
the next step.
3
Shows the current
weight value
tESt <-> 0.0 To zero the display, press the Kg-Lb key.
To advance to the next step, press the
REWEIGH key.
4
Shows the current
A/D count
A-d <-> XXXX To advance to the next step, press the
REWEIGH key.
5
Checks for current
battery level
bAt <-> XXX or nO bAt If the nO bAt is displayed, there are not
batteries in the unit, or the unit is
operating on its external AC adaptor.
To advance to the next step, press the
REWEIGH key.
6
This shows the
calibration factor.
FACtOr <-> XXXXX Press the REWEIGH key to return to the
top of the test menu.
To exit test mode, press the REWEIGH
and Kg-Lb key simultaneously, or turn off
the indicator.
Table 8-2. Test Mode Menu