Troubleshooting guide
Service Information 23
8.2 Test Mode
The test mode menu is a special mode used for checking four very important parameters which are useful in 
knowing the system’s state and for troubleshooting. Entry into this mode is possible only when the scale is turned 
off. Turn the scale on. While 
Start is displayed, press and hold the REWEIGH key until both the middle arrows 
appear on the display. 
The test mode has four parameters. They are:
•Weighing
• Internal count
• Battery indication
• Calibration factor
Alternating between the parameters is performed by pressing the REWEIGH key.
Press the Kg-Lb key to zero the scale in test mode.
Press the Kg-Lb + REWEIGH to exit weighing mode. 
Figure 8-1. Test Mode Menu Structure
The display shows a 
STOP message
The load on the scale exceeds the 
capacity of the scale 
Remove the excess weight and use the scale 
according to manufacture’s specs.
The display shows 
LO Bat message
The battery is low Recharge the battery.
The display shows 
Err message as 
detailed in the table 
below
Err 2 Low saturation state (low A/D) The load cell is not connected properly. 
Check the cables and mechanical 
connections. If the problem persists, replace 
the set of load cells.
Err 3 High saturation state (high A/D) See Err 2
Err 6 Unstable weight. Cannot calibrate Check the load cells’ mechanical 
surroundings and see that nothing touches 
them and that the cables are properly 
welded.
Err 7 Mathematical error; division by zero. 
Cannot calculate calibration factor.
This error will show when trying to calibrate 
the unit with no calibration weight on the unit.