Troubleshooting guide

Service Information 23
8.2 Test Mode
The test mode menu is a special mode used for checking four very important parameters which are useful in
knowing the system’s state and for troubleshooting. Entry into this mode is possible only when the scale is turned
off. Turn the scale on. While
Start is displayed, press and hold the REWEIGH key until both the middle arrows
appear on the display.
The test mode has four parameters. They are:
•Weighing
Internal count
Battery indication
Calibration factor
Alternating between the parameters is performed by pressing the REWEIGH key.
Press the Kg-Lb key to zero the scale in test mode.
Press the Kg-Lb + REWEIGH to exit weighing mode.
Figure 8-1. Test Mode Menu Structure
The display shows a
STOP message
The load on the scale exceeds the
capacity of the scale
Remove the excess weight and use the scale
according to manufacture’s specs.
The display shows
LO Bat message
The battery is low Recharge the battery.
The display shows
Err message as
detailed in the table
below
Err 2 Low saturation state (low A/D) The load cell is not connected properly.
Check the cables and mechanical
connections. If the problem persists, replace
the set of load cells.
Err 3 High saturation state (high A/D) See Err 2
Err 6 Unstable weight. Cannot calibrate Check the load cells’ mechanical
surroundings and see that nothing touches
them and that the cables are properly
welded.
Err 7 Mathematical error; division by zero.
Cannot calculate calibration factor.
This error will show when trying to calibrate
the unit with no calibration weight on the unit.
Problem Symptom Description Solution
Table 8-1. Troubleshooting Table