Troubleshooting guide
14 Summit SB-1000 Bariatric Wheelchair Scale Installation Manual
5.0 Scale Calibration
Before you can calibrate the scale, verify and set all scale parameters which are noted in “Indicator Installation 
and Configuration” on page 7.
Remove the configuration/calibration jumper (see page 8). Turn the scale on. While StArt is displayed, press and 
hold both the 
Kg-Lb key and the REWEIGH keys until CAL is displayed. 
To do a calibration on the scale, the scale must be turned off. 
NOTE: The calibration weight must be no less than 60 lb (28 kg) and no more than 300 lb (135 kg).
The following chart illustrates the calibration procedure
Step Function Display Available Parameters
1 Turn the scale on. 
While START is 
displayed, press and 
hold the 
Kg-Lb and 
REWEIGH keys until 
CAL is displayed. 
Enters into calibration 
mode
StArt The scale automatically advances to 
Step 2. 
2 Cal mode entered CAL To advance to the next step, press the 
REWEIGH key.
3 Sets the value of the 
calibration weight 
you are going to use 
for calibrating the 
span value of the 
scale.
LOAd <-> XXX.X Use the numeric data entry (explained in 
“Numeric Data Entry” on page 9) to set a 
calibration weight.
To advance to the next step, press the 
REWEIGH key.
4   CLEAr Clear the platform and be sure of the 
scale’s stability.
To advance to the next step, press the 
REWEIGH key.
5 Calibrate Zero - - - - - 
PUT<->XXXX
Place the requested weight on the scale. 
This will display for a few seconds.
To advance to the next step, press the 
REWEIGH key.
6 Calibrate Span CAL 
FAC tOr <-> X.XXX
This will be displayed for a few seconds 
and shows the current calibration factor. 
To advance to the next step, press the 
REWEIGH key.
7 SAvE To save the new calibration, press the 
REWEIGH key.
To exit without saving the calibration 
press the ZERO key.
8 dONE The scale displays that it has saved that 
calibration value and automatically 
advances to the last step.
9 StArt The scale reboots.
10 Con En Replace the configuration jumper.
Table 5-1. Calibration Menu