4.4 Reset To Factory Defaults
The configuration parameters may be reset to factory defaults while in the configuration mode. To do so, enter 
the configuration mode. Press the 
REWEIGH key once to advance to displaying the software version (ID <-> 
11305). Press and hold the 
BMI key until =DEF= is displayed. At this point, if you wish to perform the default 
function, press the 
REWEIGH key and the scale will show DONE, then it will restart. If you do not want to perform 
the default function, press the 
ON/OFF key.
After performing the default function, the scale will require re-calibration. In addition, the scale will be set 
up to a capacity of 600 lbs. The capacity varies by model - refer to the serial label on your scale and 
reconfigure the capacity (FULL) and display divisions (rOUnd) as necessary in the configuration menu. 
15 Set Option defaults Set < - > 1 To change, press the Kg-Lb key.
0 - default
Options = (OPx)
1 = US default (US defaults are 
OP1-1, OP2-0, OP3-0, OP4-0, OP5-0, 
OP6-1, OP-7-0, OP8-1, OP9-1, 
OP10-1).
2 = European 1 default
3 = European 2 default
9 = Used if OP1 through OP10 have 
been changed to something other than 
the factory default settings.
Note: Always set to 1 or 9 for US 
indicators. To advance to the next 
step, press the REWEIGH key.
16 Determines the 
automatic shut off 
time when the scale is 
not in use. Options are 
between one and 20 
minutes.
This is used when the 
unit is battery 
operated.
t-OFF <-> 5 Press and hold the Kg-Lb key to scroll 
through the furthest right hand digit on 
the display (0-9). If you want a value 
from 10 to 19, with the display sitting 
at 1, double press the 
Kg-Lb key and 
10 appears. Press and hold the Kg-Lb 
key to begin scrolling 10-19. If you 
want a value of 20, with the display 
sitting at 2, double press the 
Kg-Lb 
key and 20 will appear.
To advance to the next step, press the 
REWEIGH key.
17 StArt or SAvE If no changes were made in the 
parameters the display will 
automatically show 
StArt and then 
returns to normal weighing mode.
If a parameter was changed, the 
display will show 
SAvE. To sa v e  
changes made, press the REWEIGH 
key. To return to weigh mode without 
saving changes, press the 
Kg-Lb key.
18 Con En Replace the configuration jumper.
Step Function Display Available Parameters
Table 4-2. Configuration Mode Menu