Troubleshooting guide

8 Summit SB-1000 Bariatric Wheelchair Scale Installation Manual
Load Cell Connections
To gain access to the load cell connection point, remove the four back retaining screws as shown in the left hand
side of
Figure 4-3. Figure 4-2 illustrates where the load cell connection point is.
Figure 4-2. Load Cell Connection Point
4.2 Scale Configuration
Options and parameter setup are done through the scale configuration section and is used for setting values and
various parameters and options that are essential for the functioning of the system. Entry into this mode is
possible only when the scale is turned off.
4.3 Enabling Configuration or Calibration Modes
Before the scale will enter either the Configuration or Calibration mode, the configuration enable jumper must be
removed.
Access to that jumper is gained by removing the back cover of the indicator. Remove the four back cover
retaining screws as shown on the left side of
Figure 4-3. With the cover removed, the jumper can be seen sticking
through the hole in the rear housing (shown in Figure 4-3 - right side). Remove that jumper to gain access to the
configuration and calibration modes.
Figure 4-3. Gain Access to the Configuration/Calibration Enable Jumper on the Back of the Indicator
After configuration or calibration is done, the jumper must be replaced for normal scale operation.
NOTE: A display of Con En indicates that the jumper is not in place. Put jumper on both pins to return the scale to
normal weigh mode.
Scale connection (shown in
Figure 4-2 with the plastic
cover removed)
Remove retaining screws x 4 (only 2 showing)
Remove jumper to gain access to configuration
and calibration modes