Specifications
18 SB-1150 Installation & Operation Manual
9. Once all the digits have been entered, press the ENTER key and Clear is displayed.
10. Make sure the scale platform is clear of weight and press the ENTER key again and ===== is displayed.
11. A request to put the chosen load on the platform is displayed by Put xxx.xx.
12. Put the chosen weight on the platform and press the ENTER key. ===== is displayed and then Save.
13. Press the ENTER again and the display indicates Done.
14. Press the BMI key three times to exit back out to the top level Done parameter.
15. Press the ENTER key to return to weigh mode.
To exit calibration without changing zero or span existing calibration, press the CLEAR key, then the BMI key.
8.0 Test Menu
To access the TEST menu, use the following steps.
1. Turn the scale on by pressing the On/Off key and the ENTER key simultaneously until ID flashes.
2. Press the ENTER key again.
3. Continue to press the BMI key to scroll through the various menu items.
4. Once complete, press the BMI key again and Done is displayed.
5. Press the ENTER key to start the weighing process.
Figure 8-1. Test Menu
Test M e nu
Parameter Choice Description
VER Value Displays the current software version.
BAT Value Displays the current battery level.
VALUE Value Displays the actual value.
A2D Value Displays the actual raw counts of the scale.
Table 8-1. Test Menu
Power On-Off
START REWEIGH Key
TEST
Version
BAT
A-D
FACTOR
At
Press and
hold
Power Off
Remove Config/
Calibr. Jumper